VOC Emission Test Chamber
如果您對該產品感興趣的話,可以
產品名稱: VOC Emission Test Chamber
產品型號:
產品展商: Lab Companion
產品文檔: 無相關文檔
簡單介紹
VOC Emission Test Chamber
VOC Emission Test Chamber
的詳細介紹
-
Choose from two types of test areas
-
Choose from 2 different capacities (1m³ and 2m³) of the VOC Emission Test Chamber according to the size of the specimen. You can perform VOC emission testing in various fields, from compact specimens to large specimens.
-
Supports various measurement specimens
-
This system can test everything from plywood, wallpaper, curtains, and other building materials and automotive interior parts to everyday heat-generating home appliances, such as computers and printers. In addition, it is equipped with cable ports to electrify the specimen.
-
Interactive input system for easy operation
-
The instrumentation has a color LCD screen and uses an interactive input system with touch operations to enable easy settings by anyone.
-
Ensures a TVOC value of at least 20 μg/m³ for the background concentration
-
The VOC Emission Test Chamber uses material with low organic emission and low adsorption, and it comes with a high performance clean air producing unit that removes unneeded gas in the air. It constantly provides clean air and ensures a TVOC value of less than 20 μg/m³ for the background concentration prior to test start.
-
-
VOC-1000 (air-circulating fan is provided as option)
-
-
VOC-2000
-
Accurate measurement
-
In addition to having a clear air producing unit, this system uses stainless steel in the test area and reduces the amount of volatile organic compounds (VOC) generated from non-specimen materials. The test area is mirror finished and prevents the accumulation of VOC generated from the specimen for more accurate testing.
-
Adjustable the ventilation count according to test conditions
-
The ventilation count can be adjusted in the range of 0.2 to 2.0 times/hour (0.2 to 1.5 times/hour on VOC-020 only) according to test conditions.
-
Paperless data recording
The paperless recorder enables recording of the test area temperature, humidity, and air exchange rate on a memory card (compact flash drive).
Model
|
VOC-1000
|
VOC-2000
|
Compact chamber
|
Temperature control range
|
+20 to +100°C :
At test operation (no load, with specimen)
|
+80 to +250°C :
Heat processing operation time (no load, no specimen)
|
Temperature and humidity control range
|
30 to 90%RH (at+20 to +50°C)(no load, with specimen)
|
Temperature and humidity control width
|
±0.5°C / ±5% (no load, no specimen)
|
Temperature and humidity distribution width*1
|
±1.0°C / ±5% (no load, no specimen)
|
Background concentration
|
Less than 2 µg/m³ for single substance concentration
|
Less than 20 µg/m³ for TVOC concentration
|
Air exchange rate
|
0.2 to 2.0 times per hour (at test operation)
|
0.2 to 1.5 times per hour (at test operation)
|
Air exchange rate control width
|
±5% (air exchange rate of 0.5 times per hour or higher)
|
Mass transfer coefficient
|
9 to 18m/h (water vapor conversion)
|
Air tightness
|
Air leakage under 1% of supply air amount (at air exchange rate of 0.5 times per hour)
|
Recovery rate
|
80% or higher (Toluene)
|
Effective inner dimension
(W × H × D mm)
|
1000 × 1100 × 810
|
2000 × 1100 × 810
|
(Inner capacity: Approx. 1 m³)
|
(Inner capacity: Approx. 2 m³)
|
Material (Configuration)
|
Stainless steel plate (full circumference welded structure)
|
Door effective dimension (W × H mm)
|
1000 × 1100
|
2000 × 1100
|
Door lock
|
Latch lock (6 locationss)
|
Latch lock (16 locationss)
|
Outside dimensions (W × H × D mm)
|
2250 × 1950 × 1520
|
3250 × 1950× 1520
|
Door effective dimension (W × H mm)
|
1600 × 1570
|
2600 × 1570
|
Air circulator
|
Cross line flow fan
|
Refrigerator
|
Indoor air-cooled fully sealed (Refrigerant: R134a)
|
Heater
|
Nichrome strip wire heater
|
Evaporator
|
Multistage plate fin cooler
|
Chamber weight
|
Approx. 1000 kg
|
Approx. 1800 kg
|
Accessories
|
Recorder (paperless recorder), Integrating hour meter, Caster, Sampling port for specimen collection (2 locationss), Air-circulating fan (VOC-020)
|
Utility requirements
|
Power source facilities
|
Power supply
|
AC200V 3φ 3W 50/60Hz Allowable voltage fluctuation :
Rated voltage ±5% (The allowable operating range is the rated voltage ±10%)
|
Maximum load capacity
|
11.5kVA (15.8kVA when equipped with air compressor)
|
Maximum current
|
32.0A (45.6A when equipped with air compressor)
|
Leakage breaker capacity
|
40A (50A when equipped with air compressor)
|
Compressed air
|
Discharged air flow
|
160 L/min or highe (atmospheric pressure) Oil-free
|
Control pressure
|
0.6 to 0.9MPa
|
Connection tube outer diameter
|
φ10mm
|
Exhaust
|
Exhaust duct diameter
|
φ100mm (Flange diameter φ150mm)
|
Discharge air volume
|
Max. 120 m³/h or higher
|
Construction requirements
|
-
Constructed with metal duct.
-
Total pressure loss of exhaust duct at 100 Pa or lower.
|
Supply water
|
Quality
|
Pure water (Quality equivalent to A1 water specified in JIS K0557)
|
Maximum water supply rate
|
20L (manual supply)
|
Natural drainage pipe
|
Indirect drain VP30 1 units
|
Floor load capacity
|
350kg/m²
|
350kg/m²
|
*1 Measurement method complies with Japan Testing Machinery Association (JTM). However, the calculation method is half the maximum value minus the minimum value when stable.
-
This is the measured performance at the control point after stable operation for 30 minutes with an ambient temperature of +5 to +32°C and no load.
-
The allowable voltage fluctuation is within ±5% of the rated voltage, and the allowable frequency fluctuation is within ±1% of the rated frequency.
-
TVOC concentration of installation space (supply air) must be less than 300 μg/m³, and there must not be a significant fluctuation.